[Contents]
1. Artifacts
2. Resolution Artifacts
3. Propagating Speed/Path Artifacts
4. Attenuation Artifacts
5. Doppler Artifacts
1. Artifacts
1) Any unintended information on an image that does not represent the object
2) Artifacts can be a hindrance or in some cases may be diagnostically helpful
3) Basic Assumptions
- There are several assumptions made by an ultrasound machine
- Artifacts occur when these assumptions are violated
1. Sound travels in a straight line
2. Reflections are produced by structures along the beams main axis
3. Sound travels at exactly 1540 m/s
4. Intensity of a reflection directly corresponds to a reflector’s scattering strength
5. The imaging plane is very thin
6. Sound beams travel directly to a reflector and back to source
2. Resolution Artifacts
1) Axial resolution artifact [★]
- Closely spaced targets of varying distances
- Axial Resolution
: Ability to differentiate between two objects along the long axis of the ultrasound beam
= SPL/2
- Axial Resolution Artifacts appear when all 3 conditions occur:
1. Two or more reflectors are closer together than SPL/2
2. Only one reflector will appear on the image
3. Reflectors are parallel to the beam axis
- Hindrance
: Produce fewer reflectors on the image
: Actual anatomic data is missing
- Prevention
: by using higher frequency transducers with short distinct pulses
2) Lateral Resolution Artifact [★★]
- Measuring lateral width of a target on an ultrasound phantom
- Lateral Resolution
: Ability to differentiate between two objects that lie perpendicular to the ultrasound beam
= beam width
- Lateral Resolution Artifacts appear when all 3 conditions occur:
1. Two or more reflectors are closer together than the width of the beam
2. Only one reflector will appear on the image
3. Reflectors are perpendicular to the beam axis
- Hindrance
: Produce fewer reflectors on the image
: Actual anatomic data is missing
- Prevention
: by using narrower beams
3) Elevational Resolution Artifact (= Slice-Thickness A., partial volume artifact) [★★★★★]
-
- Due to beam width perpendicular to the scan plane.
- Affect imaging quality by displaying anatomic structures (reflectors) in the incorrect imaging plane
- These reflections can cause hollow structures to fill in
- Hindrance
: Causes anechoic structures to have low level echoes or false debris
- Example: false appearance of debris in simple cystic structure
- Prevention
1. Turn on Harmonics
: sound beam in this mode is narrower than in regular gray scale mode
2. Disc Shaped Elements
: Provide the thinnest slices and the best elevational resolution.
3. Newer Transducers
- 1.5-D arrays (multirow array transducer) [★★]
· They create thinnest beams with improved slice-thickness
· Have multiple crystals in an up and down direction
→ focus the beam in the thickness plane
· exhibit least amount of volume averaging
4) Contrast Resolution Artifact
- Inability of a gray-scale display to distinguish between echoes of slightly different intensities
: How many Shades of gray can be displayed
- Factoring Components
: With a decreasing number of bits per pixel, less shades of gray appear
- Less shades of gray = worse contrast resolution
- Cause: Not enough bits per pixel in image memory
- Poor contrast resolution
: Image appears more black-and-white, with few shades of gray in between
: Showing less detail
- Hindrance
: Fewer bits per pixel = fewer shades of gray = degraded contrast resolution
- Prevention: Usage of B Color
5) Spatial Resolution Artifacts
- Spatial resolution pertains to the overall detail produced in an image
- Spatial resolution artifacts are created when display monitor fail to produce adequate image detail
- Factoring Components: Spatial resolution artifacts can be created in a number of ways
1. Pixel Density
- Low pixel density (fewer pixels per inch) → larger pixels
: Larger pixels provide blurry and less detailed images
- Higher pixel density (more pixels per inch) → smaller pixels
: Smaller pixels provide better detail, better spatial resolution
- pixel density cannot be changed
- quality may be improved by write magnification (pre-processing technique)
: sonographer chooses a region of interest (ROI) to magnify
: US rescans the image → greater number of pixels, improves SR
2. Line Density
- Low line density creates less detailed images
- controlled by US system, but can be controlled by the operator
- Modern display are equipped with more lines per page providing better SR
3. Number of horizontal lines in a display monitor.
- Monitors with less horizontal lines degrade spatial resolution
3. Propagating Speed/Path Artifacts
1) Refraction Artifact (= Lateral misregistration) [★★★★]
- bending of sound beam due to different media propagation speeds
- ex) while imaging a cyst: shadowing posterior to each lateral border of cyst
- Causes of Refraction
1. Oblique incidence
2. Difference in propagating speed on either side of the boundary
Snell’s law: sinθt/sinθi = Vt/Vi
- angle of sound transmission at an interface between media with different P.S.
- Propagating Speed (c) through the 2nd medium > 1st medium
→ Transmission angle > Incident angle
- Propagating Speed (c) through the 2nd medium < 1st medium
→ Transmission angle < Incident angle
- Hindrance
: Duplication of a reflector / shows a false structure
: Reflector misplacement
- Prevention
: Change angle, artifact shouldn’t be in same place
2) Multipath Artifacts [★]
- Artifact created when the pulse is redirected along different paths before returning to the transducer
- The transmit and returning path are not the same
- Causes: Scattering
: Random redirection of sound in many directions
: Occurs with rough surfaces
: Occurs when tissue interface is small compared to the beam
: Boundary is less or equal to the wavelength of incident beam
: independent of the direction of the incident sound
- Hindrance
: Subtle, nonspecific changes that cannot be identified on an image
: may cause abnormalities of depth or position of a structure
: result in both axial and lateral displacement of reflector
3) Mirror Image Artifact
- a second copy of a true object incorrectly appears on the opposite side of a strong reflector
- Observed in all imaging mode [★]
- Characteristics
: Replica of true reflector
: Artifact will appear deeper than true reflector
: “Mirror” is in straight line between artifact and true reflector
: Artifact and true reflector are equidistant from mirror
- Cause
: sound reflects off of a strong reflector and is redirected towards another structure.
Creating a mirror image of the structure on the opposite side of the strong reflector
-
: Ultrasound will show true object in correct position (arrow showing mass on liver)
: also show a mirror image of the object (hyperechoic duplicate mass below diaphragm in image)
: Artifact will be deeper and behind the strong reflector (curved line-diaphragm)
- same path as true object
: Artifact will be same distance from strong reflector as true object
- Hindrance
: could mistake mirror image for second object
: unable to correctly view area behind high reflector
- Prevention
: Change angle of incidence to vary reflectivity of interface
: Adjust focal zone or TGC at level of high reflector causing mirroring to minimize reflectivity
: Scan from multiple windows
: Use spatial compounding
(combining image information from different angles to produce a single image)
4) Comet Tail/Ringdown Artifact [★★]
- Due to merging of two closely spaced reverberations
- Similar to a reverberation without the spacing
- Associated with resonance of a gas bubble
- Causes
: Two closely placed strong reflectors parallel to beam axis
: Sound wave bounces between the two reflectors eventually returning to the Transducer
: Most common in mediums with very high propagating speeds
- Solid hyperechoic line directed inferiorly
: Unable to differentiate between individual reverberations
: Appears posterior to actual structure
: Parallel to beams axis
- Hindrance
: False reflectors are displayed
: May obscure visualization of structures posterior to reflectors
- Prevention
: Use an alternative window
- Change beam angle
: Decrease TGC in the near gain
5) Reverberation Artifact [★★]
-
- Multiple equidistant horizontal bands having decreased brightness with depth
: Only the first two are real
- Resulting in a single structure being displayed repeatedly at greater depths
- Causes
: Two strong reflectors parallel to wave axis
: Sound waves bounce between the two reflectors eventually returning to the Transducer
- Creating a longer go return-time
- Causing incorrect reflector placement on display
- Repeated hyperechoic reflections
: Equal increments of space between artifacts
: “Ladder” or “Venetian blind” appearance
- Hindrance
: False reflectors are displayed
: May obscure visualization of structures posterior to reflectors
- Prevention
: Use an alternative window
- Change beam angle
: Decrease TGC in the near gain.
* Water-path scanner
- Advantage of water-path scanner
: near-field reverberations are reduced
- Disadvantage of water-path scanner
: bubbles in the fluid can inhibit sound transmission into the body
6) Propagating Speed Error Artifact/Range Error (= Axial misregistration) [★★]
- Created when sound propagates through medium at a rate other than 1540m/s
- Displays the correct number of reflectors at incorrect depths
: Causing misplaced echoes on image
- Causes
: Error in tissue velocity/velocity calibration of system
: Sound traveling at a speed other than 1540 m/s
- Slower than 1540 m/s (ex: large mass composed primarily of fat)
: Longer go return time than machine expects
: Pulses return slowly
: System places reflections at a greater depth (overestimates distance)
- Faster than 1540 m/s
: Shorter go return time than machine expects
: Pulses return very quickly
: System places reflectors at a shallower depth (underestimates distance)
- Displaces true reflections
- Helpful
: Conveys important information on the image
- Can provide tissue texture
- Hindrance
: Inaccurate placement may appear like pathology
- Need to look at in other views to confirm anatomy
- Prevention
: Currently cannot be prevented
: Use alternate viewing window
- Change beam angle
7) Focal Banding/Focal Enhancement Artifact
-
- Special form of enhancement
: side to side region of an image appears brighter (hyperechoic) than tissues at other depths
- Occurs in the focal region of the transducer when using multiple focal zones
- Cause
: increased intensity due to multiple foci
- Structures at the focus appear brighter than those at other depths
: An entire horizontal region (band) of tissue appears hyperechoic
- Results from increased intensity at the focus
: same appearance as incorrect TGC setting
- Hindrance
: The brightening of echoes around the focus (intensity increased by narrowing of the beam)
: Higher intensity causes a hyperechoic horizontal band across the display
→ can be mistaken for a mass
- Prevention
: Decrease the number of foci
: Change the location of multiple foci
4. Attenuation Artifacts
1) Shadowing Artifacts [★]
-
- The weakening of echoes distal to a strongly attenuating or reflecting structure [★★]
Or from the edges of a refracting structure
- Causes: result of too much attenuation
: A strongly attenuating or reflecting structure weakens the sound distal to it (attenuation)
→ echoes from the distal region are weak and appear less echogenic (like a shadow)
- Hypoechoic/Anechoic area parallel to sound beam
- Hindrance
: May hide or prevent visualization of a deeper structure
: difficult to obtain information about objects in the far field or within the shadow artifact
: Prevents visualization of true anatomy on the scan, resulting in missed information
- Helpful
: May provide valuable diagnostic information
- helps to characterize tissue
- ex) calcified plaques, stiff breast lesions, and stones
- Prevention
: Image structure in several angles to avoid missing information
: use tissue harmonic imaging (which produces thinner beam reducing slice thickness) [#1]
- Shadowing may not be displayed if beam width is greater than calcification
: due to volume averaging
: To display acoustic shadow
→ beam width ↓ (frequency ↑ and/or improve focusing)
2) Enhancement Artifacts [★★★]
-
- The strengthening of echoes from reflections that lie behind a weakly attenuating structure
- A hyperechoic region that extends beneath structures with abnormally low attenuation
- Opposite of shadowing
- Causes
: Decreased attenuation through a fluid-filled structure
: Tissues with low attenuation (ex: hematomas and abscesses)
- Helpful
: Provide valuable diagnostic information helping to characterize tissue
- Hinder
: Cause blockage and prevent from seeing something important
- Prevention
: Reduced with spatial compounding
: Several directional approaches allow the beam to get around the attenuating structure
5. Doppler Artifacts
1) Aliasing [★★]
- High velocities appear negative
- With PW doppler, high velocity measurements are inaccurate
: if the pulsed doppler sampling rate (PRF) is too low in comparison to messed doppler shift
- Appearance of Doppler spectral information on the wrong side of the baseline
- Most common Doppler artifact
: Only occurs with PW
- Very high velocities in one direction are incorrectly displayed as going the opposite direction
- Causes
: occurs because frequency-shifted signal is sampled
(rather than recorded continuously)
: inadequately sampled shift results in aliasing
: when the Nyquist limit is surpassed
- Nyquist limit (kHz) = PRF/2 [★★]
: sampling frequency needed for detecting the doppler signal unambiguously
: when PRF cannot be increased to a level greater than 2 times doppler frequency
- Nyquist frequency: The doppler frequency at which aliasing occurs
- Peaks will be displayed on the wrong side of the baseline
- Hindrance
: Yields an incorrect direction and value
: Limited ability to correctly measure deep vessels
- Prevention
: Increase the PRF (and Nyquist frequency), which will lower the baseline
- can measure high velocities
- increased PRF may introduce range ambiguity
: zero baseline shift
: Increase the Doppler angle
: Use a lower operating frequency, which lowers the Doppler shift and shrink the spectrum
: Use a continuous wave device
- Aliasing does not occur in CW doppler [★★]
: no limit in maximum velocity
2) Ghosting
- Form of noise
- The presence of false echo signals coming from outside the main beam
- Causes
: Low frequency Doppler shifts created by slowly moving anatomy
: Slow velocity reflectors
: Need to differentiate between anatomy and moving blood cells
- ex: heart muscle, pulsating vessel walls
- Hindrance
: Gives operator false diagnosis or measurement information
: Could interpret as slow-moving blood cells
- Prevention [★★★]
: Use a wall filter to eliminate low frequency Doppler shifts
: Increase PRF – range ambiguity, less sensitivity to slow flow, reduction of color fill
* Slow flow [★★]
No color signal detected within portal vein |
- color doppler system must be sensitized to detect low frequency shifts
: slow flow produces low frequency shifts
- set PRF lower to increase sensitivity to low frequency shifts
- increasing color doppler transmit frequency
: result in larger frequency shifts from slow flow
- improve visibility and sensitivity to slow flow
: decreased penetration to flow in vessels deep within tissues
- Lower the wall filter setting
3) Cross talk (spectral mirroring) [★]
-
- Identical Doppler spectrum is shown above and below the baseline
- True flow pattern is unidirectional, but flow pattern appears bidirectional on spectrum
: Spectral Doppler appears on both top and bottom
- Causes
: Equipment malfunction (poorly designed)
: Doppler gain is set too high
: Incident angle is near 90 degrees (between sound beam and flow direction)
- Doppler shifts cannot be obtained perpendicular to the beam
: Operator error
- Hindrance
: Spectral analysis is giving a false information
- Prevention
: Change angle of the transducer
: Apply gel (coupling medium)
: Fix Doppler gain
4) Twinkle artifact
- Phenomenon with unclear underlying causes that appears as a rapid alternation of color
- Immediately behind a stationary echogenic object, giving it a false appearance of movement
- Hindrance
: False appearance of movement
- Helpful
: Useful in detection of certain clinical conditions
- Ex: Urolithiasis (twinkle sign → positive reading that a stone is present)
: Can enhance accuracy and sensitivity
- Prevention
: Not having very strong scatters
: Eliminating the phase detection process
: Unfortunately, the operator has no control over these things
5) Flash artifact
- high-amplitude, low-frequency shift signal
- reduced by increasing wall filter
: reduces sensitivity to low frequency shifts ass/w slow flow
Reference
* Davies Ultrasound Physics review
* https://sites.google.com/site/lindadmsportfolio/ultrasound-physics/
* https://sites.google.com/site/nataljasultrasoundphysics/
* https://sites.google.com/site/ektasphysicseportfolio/doppler
'Ultrasonography' 카테고리의 다른 글
[ARDMS] 미국초음파자격시험에 도전해보기로 했다 (4) | 2018.09.21 |
---|---|
[Ultrasound Physics] OTHERS important for SPI ARDMS (1) | 2018.09.03 |
[Ultrasound Physics] RESOLUTION (0) | 2018.09.03 |
[Ultrasound Physics] REFLECTORS (0) | 2018.09.03 |
[Ultrasound Physics] IMAGE PROCESSING (0) | 2018.09.02 |